Abstract:With the widespread application of electronic medical equipment in hospitals, the threat of leakage current to patients is becoming greater and greater, especially in places where life is critical. The patients are directly exposed to various electrodes and sensors during surgery or anesthesia. Inserted into the human body, even the slight leakage current may cause the patient to be electrocuted. In addition, some medical equipment is used to maintain the lives of critically ill patients. Once the equipment is powered off, it will also threaten the lives of patients. Therefore, for the electrical design of the special place of the hospital, the IT system should be used for power supply in strict accordance with national standards and regulations.This article introduces the application of Medical isolated power supply in Pitsanulok hospital project.


Keywords:Medical IT,Insulation monitor device

 

0. Project Overview

Pitsanulok hospital project is located in Thailand,the project is mainly used for Pitsanulok hospital medical equipment update .

 

1. Background

Electric shock and system fault caused by power distribution system are mostly caused by the grounding fault of equipment line. Grounding fault will directly bring the harm of power supply interruption; when the earth fault leakage is serious, it will even burn electrical equipment or lines, causing irreparable loss.

 

In Medical places, especially those life-threatening class 2 medical places, patients in surgery or anesthesia state, without self-behavior ability, small leakage current can lead to electric shock death of patients. Therefore, the electrical design of this special place should be carried out strictly in accordance with the relevant national standards and specifications.

 

 图片 109


IT power distribution system (also known as isolated power supply system or ungrounded system) has very small leakage current to ground due to its neutral point ungrounded. When the first insulation fault occurs, the power supply of the equipment will not be affected. So the system can continue to operate. The insulation monitoring device installed in the system can detect the fault in time and send out the corresponding fault warning, prompting the relevant personnel to remove the fault at the appropriate time. For the operation room, intensive care unit and other medical places with high requirements for power supply continuity and safety, local IT system power supply can ensure the security and reliability of power supply for important loads. 图片 99

 

2. Description of five sets of medical IT productsIntroduction

 Medical IT system insulation monitoring and fault locating products (five-piece set) include AITR series medical isolation transformer, AIM-M100 medical intelligent insulation monitoring instrument, AKH-0.66P26 current transformer, ACLP10-24 DC power supply, and AID series(AID120/AID150) centralized alarm and display instrument, which are shown in Table 1.

Table 1 Medical IT System Insulation Monitoring Products

Name and Model

    Product Pictures

Descriptions

AITR series medical isolation transformer

 QQ截图20220113214458

    

AITR series isolation transformer is specially used in medical IT system, and the core superposition adopts the silicon steel sheet imported from Japan, which has very small losses. The windings are treated with double insulation and have electrostatic shielding layer, which reduces electromagnetic interference between windings. The PT100 temperature sensor is installed in the windings to monitor the temperature of transformer. The whole body is treated with vacuum invasion paint, which increases mechanical strength and corrosion resistance. The product has good temperature rise performance and very low noise.

AIM-M100 medical insulation monitoring device

 QQ截图20220113214716


AIM-M100 medical insulation monitoring device is special for medical IT system.It is used to monitor the insulation status of IT system.It can send out alarm signals when insulation fault occurs.

AKH-0.66P26 current transformer

 图片 102



The AKH-0.66P26 type current transformer is the protective current transformer which the maximum measurable current is 60A and the transformation ratio is 2000:1. The current transformer is directly fixed inside cabinet by screwing

ACLP10-24 DC power supply

 QQ截图20220113214823

The DC voltage regulator module for exclusive use of instrument employs completely isolated linear transformer with the characteristics of stable output voltage, small ripple and high-grade withstand voltage etc. It is also equipped with the electric indicative function of power on. The module adopts the installation method of standard guide rail which can be installed on the same guide rail as the insulation monitoring device.

AID series external alarm and displayer

AID120

 QQ截图20220113214844

It is fit for embedding into operating room or nurse station internal wall installation which can monitor one set of insulation monitoring device with the function of sound-light alarm on failures like insulation, overload, overheat and equipment failure etc.

AID150

 QQ截图20220113214902





The concentrated alarms and displayers which can do centralized monitoring of the data of 16 sets of medical  insulation monitoring devices to the maximum and are capable of remote sound-light alarm. AID150 can also monitor the data of multiple sets of AIM-R100 residual current monitoring devices.

 

3.现场图片 On site Photos

 

QQ截图20220113214931

QQ截图20220113215454 

 